RK2675YM series medical leakage current tester
Product Introduction
The ordinary leakage current tester, also known as the contact current tester, is used to measure the leakage of electrical appliances that is not related to work caused by insulation or distributed parameter impedance of the working power supply (or other power sources). Its input impedance simulates the impedance of the human body and meets the requirements of GB4706.1 and GB9706.1-2020. The RK2675YM series medical leak tester products are designed in accordance with international and domestic safety standards such as IEC, ISO, BS, UL, JIS, etc. Leakage current alarm values can be continuously and arbitrarily set according to different safety standards and user needs;
application area
1. Medical equipment: various types of new medical instruments and medical supporting instruments, cardiac monitoring, medical imaging, biochemical analysis instruments, blood pressure monitors and thermometers, and other home medical equipment
2. Diagnostic and therapeutic equipment: X-ray diagnostic examination equipment, ultrasound diagnosis, nuclear medicine, endoscopic system, facial features treatment equipment, dynamic analysis treatment equipment, low-temperature freezing equipment, dialysis treatment equipment, emergency equipment
3. Ward nursing equipment and appliances: various types of beds, cabinets, operating chairs, beds, etc
4. Auxiliary equipment: medical nursing data and image processing equipment, rehabilitation equipment, disability specific equipment, etc
5. Oral medical equipment: dental diagnostic medical equipment, dental surgical instruments, dental technician equipment, medical magnetic resonance equipment
Performance characteristics
1. Test circuit (MD). Range change unit
2. It is convenient for users to choose the appropriate range according to the actual load size
3. The display module displays the test voltage, actual leakage current, and test time
4. The over limit alarm circuit completes the alarm and indication of non standard products and automatically cuts off the output power supply
5. The experimental voltage adjustment device can adjust the appropriate test voltage according to different standards
Model | RK2675YM | RK2675YM-1 | RK2675YM-2 | RK2675YM-3 | RK2675YM-5 |
Test Voltage | 0~300V | ||||
Test Current | AC/DC:0~200μA AC/DC:0.2~2mA AC:2~10mA | ||||
Test Accuracy | ±5% | ||||
Test Time | 0~99s (Adjustable Continuously) | ||||
Transformer Capacity | 500VA | 1000VA | 2000VA | 3000VA | 5000VA |
Output Waveform | Sine Wave | ||||
Power requirements | 220V ±10%50Hz 2% | ||||
Working environment | 0℃~40℃ ≤85%RH | ||||
Exterior Dimension(mm) | 375*280*200mm | 375*280*200mm | 430X380X200mm | 430X380X200mm | 505X470X270mm |
Weight | 14Kg | 19.6Kg | 34.8Kg | 41.5Kg | 73.2Kg |
Accessory | Power Line,Alligator Clip |