RK9914/RK9914A/RK9914B/RK9914C/RK9915/RK9915A/RK9915B Program controlled AC / DC withstand voltage tester

RK9914:AC (0.00~5.00)kV DC(0.00~6.00)kV AC 0~100mA;DC:0~50mA

IR:0.10-5.00kv 100GΩ
RK9914A:AC (0.00~5.00)kV DC(0.00~6.00)kV AC 0~100mA;DC:0~50mA
RK9914B:AC (0.00~5.00)kV AC 0~100mA
RK9914C:AC (0.00~5.00)kV DC(0.00~6.00)kV AC 0~50mA;DC:0~25mA


Description

Parameter

Accessories

Video

RK9914A/B/C  program-controlled withstand voltage tester

 

Product introduction

 

This series of program-controlled withstand voltage tester adopts high-performance safety gauge tester designed by high-speed MCU and large-scale digital circuit, and its output voltage

 

The rise and fall of output voltage and the frequency of output voltage are completely controlled by MCU, which can display the breakdown current and voltage value in real time,

 

It also has software calibration function and is equipped with PLC, RS232C, RS485, USB and LAN interfaces to form a comprehensive test system with computer or PLC system.

 

It can quickly and accurately measure the safety regulations of household appliances, instruments and meters, lighting appliances, electric heating tools, computers and information machines.

 

Applicable standards: IEC60335-1, GB4706 1. Ul60335-1 safety of household and similar electrical appliances Part 1: General requirements ul60950, gb4943, iec60950

 

Safety requirements for information technology equipment ul60065, gb8898, IEC60065 audio, video and similar electronic machinery iec61010, gb4793 1 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 1: General requirements

 

application area

 

Components: diode, triode, high voltage silicon stack, various electronic transformers, connectors, high voltage capacitors

 

Household appliances: TV, refrigerator, air conditioner, washing machine, dehumidifier, electric blanket, charger, etc

Insulating materials: heat shrinkable sleeve, capacitor film, high voltage sleeve, insulating paper, insulating gloves, etc

Electric heating and electric tools, instruments and meters, etc
Performance characteristics

 

7-inch TFT (800 * 480) is used to display the setting parameters and test parameters, with eye-catching and rich display content, high current and high power

 

Software upgrade via USB interface

 

Adjustable high-voltage rise and fall time, which can meet the requirements of different test objects

Test results can be saved synchronously

The user-friendly operation interface supports direct input of digital keys, and dial input and operation are simpler

Bilingual operation interface in Chinese and English to meet the needs of different users

Minimum DC current resolution 0.001 μ A

Standard PLC interface, RS232 interface, RS485 interface and USB interface


  • Previous:
  • Next:

  • 英_03

    parameter model RK9915 RK9915A RK9915B
    ACW Output voltage range (0.05 ~ 5.00)kV
    Maximum output power 1000VA(5.0kV 200mA)
    Maximum rated current 200mA
    Output waveform Sine Wave DDS+ Amplifier
    DCW Output voltage range (0.05 ~ 6.00)kV /
    Maximum output power 600VA(6.0kV 100mA) /
    Maximum rated current 100mA /
    IR Output voltage (DC) (0.10 ~ 5.00)kV / /
    Resistance test range ≥500V 0.10MΩ-1.0GΩ±5%
    1.0G-50.0GΩ ±10%
    50.0GΩ-100.0GΩ ±15%
    <500V 0.10MΩ-1.0GΩ ±10%
    1.0GΩ-10.0GΩ ±15%
    / /
    Voltmeter Range AC(0.05 ~ 5.00)kV DC(0.05~6.00)kV AC(0.05~5.00)kV
    Accuracy ± (3%+5 words)
    Setting error ± (3%+5 words)
    Ammeter Measurement range AC: 0~200mA DC: 0~100mA AC: 0~200mA
    Measurement accuracy ± (3%+5 words)
    Timer Range 0.0-999.9S
    Minimum resolution 0.1S
    Test time 0.1S-999S OFF= continuous test
    Arc detection 0-20mA
    Output frequency 50Hz/60Hz
    Operating temperature 0-40℃ ≤75%RH
    Power requirements 110/220 ±10% 50Hz/60Hz ±3Hz
    Interface RS232, USB, PLC, LAN, RS485
    Screen 7-inch TFT 800*480
    Dimensions (D × H × W) 670*245*440mm
    Weight 63KG
    Random standard accessories Power cord RK00004, RS232 communication cable RK00002, RS232 to USB cable RK00003, USB to square port cable RK00006, RK26003A test cable, instruction manual (electronic version), RK00048 test cable, host computer (official website download), RK8N+ high voltage stick
    Optional accessories RK00031 USB to RS485 Female serial port cable industrial grade connection cable 1.5 meters long
    Contact check (resistance software judgment) Optional open or close