The Programmable Leakage Current Tester Is A 5-Inch TFT LCD Display. The Tester Adopts 32-Bit High-Speed MCU And Large-Scale Digital Circuit Design,Built In GB9706.1-2007 (IEC60601-1:1998) Body Impedance Simulation Network,The Card MD Network Interface Contains MD-A (In Line With GB/T12113-2003 And GB4793.1-2007)、MD-B(Accord With GB/T12113-2003、GB4793.1-2007、GB4706.1-2005、GB4943.1-2011、GB8898-2011、GB7000.1-2015)、MD-C(Accord With GB/T12113-2003、GB7000.1-2015)、MD-D(Accord With GB4793.1-2007)、MD-E(Accord With GB4943.1-2011、GB4793.1-2007)、MD-F(Accord With GB7000.1-2015)。Load Voltage And Current Overrun Protection; MD Network Multiple Fast Protection, Safe And Reliable. Dynamic And Static Power Supply Test States; Leakage Current Test Automatically Switches L (Phase Line) And N (Zero Line). Its Setting And Operation Are Very Simple, And Provides PLC Remote Control Interface, RS232C, RS485, USB And Other Interfaces, Which Is Convenient For Users To Quickly Combine Into A Comprehensive Test System.
Packing&Shipping
for reference .Then make payment as the way you like , as soon as the payment have been confirmed ,we will arrange shippment
within 3 days .
been confirmed.
1、Specific Parameters | |||||
Model | RK9950(Passive) | RK9950A(500VA) | RK9950B(1000VA) | RK9950C | |
Basic Functions | Screen Size | 5-Inch TFT LCD Screen | |||
Number Key | Parameter Setting Digital Input | ||||
Coding Switch | Parameter Selection And Confirmation Function | ||||
Up, Down, Left And Right Function Keys |
Parameter Setting Up And Down Selection Function | ||||
LOCK Keyboard Locking Function | Prevent Accidental Modification Of Test Conditions Or Prohibit Modification Of Test Conditions | ||||
Alarm Function | Voice Alarm | ||||
Communication Interface | RS232C、RS485、USB | ||||
USB Interface | Copy, Copy And Storage Functions | ||||
Control Interface | HANDLER(PLC) | ||||
Test Method | Dynamic, Static | ||||
Load Voltage(AC) | Range | 0-300V | |||
Accuracy | + (2% ×Display Value +0.5V) | ||||
Frequency | 50/60Hz | ||||
Load Current(AC) | Maximum | 25A | |||
Over Current Protection | Sound And Light Alarm, Cut Off The Load Output | ||||
Leakage Current Setting | Upper Limit Setting | ||||
Range | 0.001-20.00mA | ||||
Resolution | <10mA,0.001/Step >10mA, 0.01mA/Step | ||||
Lower Limit Setting | |||||
Range | 0.000-20.00mA | ||||
Resolution | <10mA,0.001/Step >10mA, 0.01mA/Step | ||||
Explanation | The Lower Limit Is Set To 0 Without Judgment | ||||
Leakage Current Measurement | Range And Accuracy | 0.001-0.050mA DC-10KHz | ± (5% × Display Value + 5 Words) | ||
0.050-20.00mA DC-10KHz | ± (2% × Display Value + 2 Words) | ||||
0.050-20.00mA 10KHz-1MHz | ± (5% × Display Value + 5 Words) | ||||
MD Simulated Human Network | 8 Standard Networks | MD-A(GB/T12113-2003、GB4793.1-2007)、MD-B B1(GB/T12113-2003、GB4793.1-2007、GB4706.1-2005、GB4943.1-2011、GB8898-2011、GB7000.1-2015)、MD-C(GB/T12113-2003、GB7000.1-2015)、MD-D(GB4793.1-2007)、MD-E(GB4943.1-2011、GB4793.1-2007)、MD-F(IEC60598-1:2014、GB4793.1-2007)、MD-G(GB4943.1-2011、IEC60950-1:2005、GB4793.1-2007、IEC61010-1:2001) | |||
Test Time | Range | Single Timing Test: 0.1-999s ± 1%; Closing Time Is Continuous Test | |||
Explanation | When Setting 0S Test, Only N-Wire Is Tested,No L Conversion | ||||
Output Power | Passive | 500VA | 1000VA | Three Phase Passive | |
External Power Supply | External Connection | Built In | Built In | External Connection | |
2、General Technical Index | |||||
General Technical Index | |||||
Working Temperature And Humidity | 0℃-40℃,≦75%RH | ||||
Power Supply | 100V-121V,198V-242V,47.5-63Hz 5A 250V | ||||
Fuse Size | 5A 250V 10A 110V | ||||
External Volume(D×H×W) | 430mm×105mm×350mm | ||||
Weight | 13KG | 16KG | 18KG | 14KG |
MODEL | PICTURE | TYPE | SUMMARY |
RK995001 | Standard | The Instrument Is Equipped With Standard Power Cable, Which Can Be Purchased Separately. | |
RK20K | Standard | The Instrument Is Equipped With Standard Data Cable, Which Can Be Purchased Separately. | |
RK00001 | Standard | The Instrument Is Equipped With Standard Power Cord, Which Can Be Purchased Separately. | |
Certificate Warranty Card | Standard | Instrument Standard Certificate And Warranty Card. | |
Factory Calibration Certificate | Standard | Calibration Certificate Of Instrument Standard Products. | |
Instructions | Standard | The Operation Manual Of The Instrument Standard Product. | |
PC software |
Optional
|
The instrument is equipped with 16g U disk (including upper computer software). | |