Ungakuqhuba kanjani ukuhlolwa kokuphepha kumandla angaphakathi wokuhlinzeka ngezinto zokwelashwa? Ngamandla angaphakathi ahlinzekwa ngezinto zokuhlola zezokwelashwa, azikho ucingo oluyisisekelo, ngakho-ke asikho isidingo sokwenza ukuhlolwa okuyisisekelo. Okokuqala, kukhona i-voltage volustage test. Kulesi sivivinyo, sasebenzisaRK267200I-Medical Voltage Volstage I-Aster I-Port ixhume isibhamu esiphakeme samandla we-wiring port yangaphandle bese ixhuma enye indawo yokuzithandela ye-wiring yesibhamu esiphakeme se-voli.
I-terminal yokubuya yamanje yensimbi ixhumeke kwi-Shell yento okufanele ihlolwe. Ngemuva kokuthi i-wiring isiqediwe, vula amandla kagesi. Okokuqala khetha ibanga. Ithuluzi line-2MA kanye namabanga angama-20. Ngemuva kokuthi uhla lukhethiwe, lungisa umkhawulo ophezulu wesamanje. Cindezela inkinobho esethiwe ebusweni bethuluzi, thatha i-screwdriver efihliwe ukuze ulungise amanje sokulungiswa kwe-potentiometer yamanje. Jikelezisa iwashi lamanje lokukhulisa, kanye nokuqina kwe-counterclock Ngemuva kokulungisa i-Preset Yamanje, cindezela inkinobho esethiwe, hlela isibhamu esiphakeme samandla we-voltage ngamandla angaphakathi, cindezela inkinobho yokuqalisa ethekwini lokushaja lento yokuhlola yezokwelapha, lungisa i-voltage edingekayo, bese ukhipha inkinobho yokushintsha ukuze uqedele isilinganiso.
Okulandelayo, ukukala ukuvuza kwamanje futhi usebenzise uMerrick RK7505Y ukuvivinya ukuvuza okuvuzayo kwesiguli kwemishini yokuhlinzekwa kwamandla wangaphakathi. Mane nje uxhume isiphetho esiphakeme se-MD engxenyeni yesicelo kanye nokuphela okuphansi kwe-MD kuya kwelinye iphuzu engxenyeni yesicelo, futhi ezimbili azikwazi ukuxhunywa.
Ngemuva kokuthi i-wiring isiqediwe, vula ukuhlinzekwa kwamandla, cindezela isethi ukusetha imodi ku-5, ukhethe uhlobo kunani oyifunayo, cindezela i-ENT ukuze ulondoloze amapharamitha, bese ucindezela inkinobho yokuqalisa eluhlaza ukuqala Hlola.
Imithethonqubo yokuphepha yezokwelapha isethi yokuphepha ezintathu isebenzisa abahloli bezokuphepha abathathu, okungukuthi i-Meiruike RK2672Mym, RK2678YM (GB9706 standard), kanye ne-RK7505Y.
Uma kukhona noma yikuphi ukushintshaniswa kwezobuchwepheshe, sicela ukhululeke ukubheka ukwesekwa kwangaphambi kokuthengisa kukaMerrick. Wamukelekile ukuza uzobuza!
Isikhathi sePosi: Aug-07-2023